User manual
70 Agilent 1260 Infinity High Performance Autosampler User Manual
5 Optimizing Performance
How to Achieve Higher Injection Volumes
How to Achieve Higher Injection Volumes
The standard configuration of the Agilent 1260 Infinity Autosampler can
inject a maximum volume of 100 µL with the standard loop capillary. To
increase the injection volume the Multidraw upgrade kit (G1313- 68711)
can be installed. With the kit you can add a maximum of 400 µL or
1400 µL to the injection volume of your injector. The total volume is then
500 µL or 1500 µL for the 1260 Infinity Autosampler with 100 µL
analytical head. Note the delay volume of your autosampler is extended
when using the extended seat capillaries from the multi- draw kit. When
calculating the delay volume of the autosampler you have to double the
volume of the extended capillaries. The system delay volume due to the
autosampler will increase accordingly.
Whenever a method is scaled down from a larger column to a smaller
column it is important that the method translation makes an allowance for
reducing the injection volume in proportion to the volume of the column to
maintain the performance of the method. This is to keep the volume of the
injection at the same percentage volume with respect to the column. This is
particular important if the injection solvent is stronger (more eluotropic)
than the starting mobile phase and any increase will affect the separation
particularly for early running peaks (low retention factor). In some cases it
is the cause of peak distortion and the general rule is to keep the injection
solvent the same or weaker than the starting gradient composition. This has
a bearing on whether, or by how much, the injection volume can be
increased and the user should check for signs of increased dispersion (wider
or more skewed peaks and reduced peak resolution) in trying to increase the
injection size. If an injection is made in a weak solvent then the volume can
probably be increased further because the effect will be to concentrate the
analyte on the head of the column at the start of the gradient. Conversely if
the injection is in a stronger solvent than the starting mobile phase then
increased injection volume will spread the band of analyte down the column
ahead of the gradient resulting in peak dispersion and loss of resolution.
Perhaps the main consideration in determining injection volume is the
diameter of the column as this will have a big impact on peak dispersion.
Peak heights can be higher on a narrow column than with a larger
injection on a wider column because there is less peak dispersion. With
2.1 mm i.d. columns typical injection volumes might range up to 5 to10 µl