Agilent 1260 Infinity High Performance Autosampler User Manual 69
Optimizing Performance
5
How to Configure the Optimum Delay Volume
The ADVR function is unlikely to be suitable for applications involving
compounds which are known to cause carry- over problems.
The best solution to reduce the delay volume is to install the 40 µL
injection upgrade kit (G4215A). The standard metering device is replaced
by a 40 µL Micro Analytical Head and a new 40 µL Loop must be
installed. To get the best results it is also recommended to order the Low
dispersion kit (G1316- 68744) and the micro flow cell for UV. This will
reduce the the delay volume by 120 µL.