User manual

Agilent 1260 Infinity High Performance Autosampler User Manual 67
Optimizing Performance
5
How to Configure the Optimum Delay Volume
How to Configure the Optimum Delay Volume
For very fast gradients over 0.5 min the delay volume of the system can
be easily reduced without changing the physical configuration of the
system. The change is achieved by changing the behavior of the
autosampler.
The 270 µL delay volume of the autosampler is due to the flow path from
the injection valve through the metering device, needle, needle seat and
connecting capillaries back to the injection valve (see Figure 16 on
page 68). To make an injection the valve switches from mainpass to
bypass so that the metering device can draw the sample into the needle
capillary. The injection is made when the valve switches back to mainpass
and the sample is flushed onto the column. The valve remains in this
position during analysis so that the autosampler is continually flushed and
hence the gradient has to flow through this delay volume to reach the
column. This can be eliminated by switching the injection valve from
mainpass to bypass after the injection has been made and the injected
sample has been flushed onto the column. In practice this can be done a
few seconds after injection and is activated by selecting the “Automatic
Delay Volume Reduction” (ADVR) function in the autosampler setup menu.
The Flush- out Factor (typically 5 times injection volume) ensures that
enough time is allowed to flush the sample out of the injector before
switching to bypass. For instance a 1 µL injection under standard
conditions effectively reduces the system delay volume by approximatly
250 µL.