58 Agilent 1260 Infinity High Performance Autosampler User Manual
4 Using the Module
Setting up the Autosampler with Agilent ChemStation
Injection Cleaning
The Injection Valve Cleaning section allows you to
specify the valve switching times at the end of overlap or
sample flush.
Times 1 ... 4 are the times when the valve switches to
bypass (for time 1) or to mainpass and bypass (for times 2,
3 and 4).The times must be specified in ascending order.
You can also switch the times to off. Between the first and
second, and second and third valve switches, a rinse is
executed using the rinse volumes specified in the Injector
Cleaning section.
Valve movements specifies the number of times that the
valve switches from mainpass to bypass at times 2, 3 and
4 in the field.The maximum value is 2; default is 1.