3 Installing the Autosampler
Flow Connections to the Autosampler
Flow Connections to the Autosampler
1 Open the front cover by pressing the button on the right side of the
module.
2 Install the capillary from the pump outlet into the port 1 of the
injection valve.
3 Install the capillary from the port 6 of the injection valve to the TCC.
Parts required Description
System
Capillaries and tubing from Accessory Kit.
Preparations Autosampler is installed in system.
NOTE
In an Agilent 1260 Infinity Liquid Chromatograph, the Autosampler is located between a
Pump (above) and the Thermostatted Column Compartment (below), see “Optimizing the
Stack Configuration” on page 35
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.
NOTE
The Autosampler can only be operated with the front and side covers closed.