User manual

Agilent 1260 Infinity High Performance Autosampler User Manual 29
Site Requirements and Specifications
2
Specifications
Local Control Agilent Instant Pilot (G4208A) B.02.11 or above
Communications Controller-area network (CAN), RS-232C,
APG Remote: ready, start, stop and
shut-down signals, optional four external
contact closures and BCD vial number
out.
Safety and maintenance Extensive support for troubleshooting and
maintenance is provided by the Instant
Pilot, Agilent Lab Advisor, and the
Chromatography Data System.
Safety-related features are leak detection,
safe leak handling, leak output signal for
shutdown of pumping system, and low
voltages in major maintenance areas.
GLP features Early maintenance feedback (EMF) for
continuous tracking of instrument usage
with user-settable limits and feedback
messages. Electronic records of
maintenance and errors.
Housing All materials recyclable.
Metering device Metering pump in high pressure flow path
1
Chromatographic conditions: Column: Agilent ZORBAX SB-C18, 2.1 x 50 mm1.8 µm
(p/n 827700-902); mobile phase: A: 0.1 % TFA in water, B: 0.1 % TFA in acetonitrile;
isocratic : %B=35 %; flow rate: 0.5 mL/min; temperature: 30 °C
2
UV-detection: Sample : 1200 ng/µL chlorhexidine (dissolved in mobile phase A), 1 µL injected and
measured on G4212A DAD (10 mm cell); Wavelength: 257 nm +/- 4 nm; ref. 360 nm +/- 16 nm; slit
4 nm, 10 Hz
3
MS-detection: Sample : 50 ng/µL chlorhexidine (dissolved in mobile phase A), 1 µL injected and
measured on Agilent 6460 QQQ (in specified conditions); MRM 1: 505.5 170 (CE: 36 V);
MRM 3: 505.5 201.2 (CE: 20 V); fragmentor: 150 V, delta EMV(+): 200 V
Tabl e 2 Performance Specifications (G1367E)
Type Specification Comment