User manual
28 Agilent 1260 Infinity High Performance Autosampler User Manual
2 Site Requirements and Specifications
Specifications
Specifications
Tabl e 2 Performance Specifications (G1367E)
Type Specification Comment
Injection range 0.1 – 100 µL in 0.1 µL increments. Up to
40 µL with reduced injection volume kit
(hardware modification required). Up to
1500 µL with multiple draw (hardware
modification required).
Precision <0.25 % from 5 – 40 µL
<0.5 % from 2 – 5 µL
<0.7 % from 1 – 2 µL
<1.5 % from 0.5 – 1 µL
Injection Accuracy 1 % (10 µL, n=10)
Pressure range Up to 600 bar (8700 psi)
Sample viscosity range 0.2 – 5 cp
Sample capacity Capacity 2 x well plates (MTP) + 10 x 2 ml
vials, 108 x 2 ml vials in 2 x 54 vial plate
plus 10 additional 2 ml vials, 30 x 6 ml
vials in 2 x 15 vial plate,
100 Micro vial tray,
plus 10 additional 2 ml vials,
54 Eppendorf tubes (0.5/1.5/2 ml) in 2 x
27 Eppendorf tube plate.
Also compatible with
the Agilent 1200 Series
sample capacity
extension for further
expansion of the
sample capacity.
Injection cycle time Typically <21 s using default conditions
and injection vomlume of 5 µL
Carry Over Typically <0.004 % For measurement
conditions see
1
,
2
,
3
Control and data evaluation Agilent ChemStation for LC
EZChrom Elite
MassHunter TOF/QTOF and QQQ
B.04.02 SP1 DSP3 or
above
3.3.2 SP2 or above
B.04.00 or above
B.03.01 SP2 or above