2 Site Requirements and Specifications
Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the
module.
Power Consideration
The module power supply has wide ranging capabilities and accepts any
line voltage in the range mentioned in Table 1 on page 27. Consequently,
there is no voltage selector in the rear of the module. There are also no
externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.
WARNING
Module is partially energized when switched off, as long as the power cord is
plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened and the module is connected to power.
➔ Make sure that it is always possible to access the power plug.
➔ Remove the power cable from the instrument before opening the cover.
➔ Do not connect the power cable to the Instrument while the covers are removed.
WARNING
Incorrect line voltage at the module
Shock hazard or damage of your instrument can result if the devices are connected
to line voltage higher than specified.
➔ Connect your module to the specified line voltage.
CAUTION
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
➔ Make sure the power connector of the instrument can be easily reached and
unplugged.
➔ Provide sufficient space behind the power socket of the instrument to unplug the
cable.