12 Hardware Information
Setting the 8-bit Configuration Switch
Communication Settings for RS-232C
The communication protocol used in the column compartment supports
only hardware handshake (CTS/RTR).
Switches 1 in down and 2 in up position define that the RS- 232C
parameters will be changed. Once the change has been completed, the
column instrument must be powered up again in order to store the values
in the non- volatile memory.
NOTE
With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed.
The preferred communication is LAN.
NOTE
The following tables represent the configuration switch settings for the modules without
on-board LAN only.
Tabl e 11 8-bit Configuration Switch (without on-board LAN)
Mode Select12345678
RS-232C 0 1 Baudrate Data
Bits
Parity
Reserved 1 0 Reserved
TEST/BOOT 1 1 RSVD SYS RSVD RSVD FC
NOTE
The LAN settings are done on the LAN Interface Card G1369B/C. Refer to the
documentation provided with the card.
Tabl e 12 Communication Settings for RS-232C Communication (without on-board LAN)
Mode
Select
12345 6 78
RS-232C 0 1 Baudrate Data Bits Parity