Agilent 1260 Infinity High Performance Autosampler User Manual 141
Maintenance
9
Replacing Peristaltic Pump Cartridge
3
Pull the cartridge forward off the motor shaft. 4 Disconnect the tubing leading to the wash port and the
tubing coming from the solvent bottle.
5 Connect the wash port tubing to the upper tubing of the
new cartridge (use sand paper to get a good grip on the
tubing).
6 Connect the tubing coming from the solvent bottle to the
lower tubing of the new cartridge.