9 Maintenance
Replacing Peristaltic Pump Cartridge
Replacing Peristaltic Pump Cartridge
When Tubing blocked or broken
Parts required # p/n Description
1 5065-4445 Peristaltic pump cartridge
NOTE
The peristaltic pump cartridge is a replaceable unit. The tubing inside the pump is not
replaceable.
1
Remove the corrugated leak tubing. 2 Press the two clips on the front of the peristaltic pump
cartridge.