User manual
132 Agilent 1260 Infinity High Performance Autosampler User Manual
9 Maintenance
Replacing the Rotor seal
Replacing the Rotor seal
When When poor injection volume reproducibility or when injection valve is leaking.
Tools required p/n Description
8710-0510 ΒΌ inch wrench
8710-2394 9/64 inch hex key
Parts required # p/n Description
1 0101-1416 Injection valve rotor seal
1 Open the front door. 2 Remove all capillaries from the Injection valve with a 1/4
inch wrench.