User manual

Agilent 1260 Infinity High Performance Autosampler User Manual 131
Maintenance
9
Exchanging the Needle Seat
3
Disconnect the seat capillary from the Injection valve. 4 With a Flat head screw driver carefully lift out the needle
seat from the holder.
Next Steps:
5 Insert the new needle seat. Press it firmly in position.
6 In the user interface exit the Change needle/seat function and exit the maintenance mode. In the Lab Advisor software
the Change needle/seat function can be found in the Tools section.