Agilent 1260 Infinity High Performance Autosampler User Manual 125
Maintenance
9
Removing the needle assembly
1
In the user interface start the maintenance mode and
select Change needle/seat function. In the Agilent Lab
Advisor software the Change needle/seat function can
be found in the Tools section.
2 Open the front door and remove the side door.
3 Turn the Needle carrier 90 ° clockwise. 4 Flip the Leak guide open.