User manual
110 Agilent 1260 Infinity High Performance Autosampler User Manual
8 Test Functions
Sample Transport Self Alignment
Sample Transport Self Alignment
The sample transport self alignment uses predefined positions on the well
plate tray to calibrate the positioning of the needle. The sample transport
self alignment is required to compensate for larger deviations in
positioning the needle carrier. The sample transport self alignment is
required after disassembling the system or when you exchange the sample
transport, the sampling unit, the tray or the MTP main board. This
function is in the calibration screen of the Lab Advisor.
When After disassembling the module or by larger deviations in the positioning of the needle.
Preparations Well plate tray needs to be installed and empty.