Agilent 1260 Infinity High Performance Autosampler User Manual 109
Test Functions
8
System Pressure Test
System Pressure Test Evaluation
System Pressure Test Failed
Probable cause Suggested actions
1 Pump leakages
Perform the Pump Head Leak test.
2 Loose or leaky fittings
Tighten the fittings or replace capillaries.
3 Autosampler leakages
Perform the Autosampler Leak test.
4 Thermostatted Column Compartment valve
leakages
Replace the TCC valve rotor seal.
NOTE
• Notice the difference between error in the test and a failed result! An error is caused by
an abnormal termination during the operation of the test, whereas a failed result
indicates that the test results were not within the specified limits.
• Often it is only a damaged blank nut (poorly shaped from over tightening) that causes
the test to fail. Before investigating any other possible sources of failure make sure that
the blank nut you are using is in a good condition and properly tightened.