User manual
104 Agilent 1260 Infinity High Performance Autosampler User Manual
7 Error Information
Module Error Messages
Vessel Stuck to Needle
Error ID: 4453
The vessel sticks to the needle when the needle moves up.
Rear Blind Seat Missing
Error ID: 4724
Rear blind seat is missing although claimed to exist by main board
information – occurs during initialization or if the blind seat location has
to be used.
Probable cause Suggested actions
1 Closing mat to rigid/thick.
Check that the closing mat is not too thick.
2 Bad X or Theta positioning and the needle
sticks into the wall between two holes.
Please contact your Agilent service
representative.
3 Defective encoder on the needle carrier
assembly.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Blind seat is missing.
Install blind seat.