102 Agilent 1260 Infinity High Performance Autosampler User Manual
7 Error Information
Module Error Messages
Invalid Vial Position
Error ID: 4042
The vial position defined in the method or sequence does not exist.
The reflection sensors on the transport assembly flex board are used to
automatically check which sample trays are installed (coding on tray). If
the vial position does not exist in the current sample tray configuration,
the error message is generated.
Probable cause Suggested actions
1 Incorrect tray installed.
Install the correct trays, or edit the method or
sequence accordingly.
2 Incorrect tray definition.
Install the correct trays, or edit the method or
sequence accordingly.
3 Incorrect vial positions defined in the
method or sequence.
Install the correct trays, or edit the method or
sequence accordingly.
4 Tray recognition defective (dirty sample tray
or defective transport assembly flex board).
• Ensure the coding surfaces of the sample
tray are clean (located at the rear of the
sample tray).
• Please contact your Agilent service
representative.