Technical information
2
Setting Up the LAN Interface
Step 4: Setting Up the LAN Interface to Use a Stored IP Address
14 Agilent 1200 Series Getting Connected Guide Thermo Scientific
Step 4: Setting Up the LAN Interface to Use a Stored IP Address
For the LAN interface to use the stored IP address that you set with a Telnet session, set the
initialization mode to use a stored IP address setting. The initialization mode is set with three
toggle switches.
Y To set the on-board LAN interface initialization mode to use a stored IP address
Set the initialization switches on the back panel of the instrument as follows:
The On position for the toggle switch is up and the Off position is down.
Y To set the Agilent G1369 A LAN card initialization mode to use a stored IP address
1. Turn off the instrument that has the LAN card and unplug the instrument from line
power.
2. Remove the LAN card (see Figure 5 on page 8) from the instrument (refer to the Agilent
G1369A LAN Interface User Manual).
3. Set the initialization switches to the initialization mode that uses a stored IP address as
follows:
The On position for the toggle switch is up and the Off position is down.
4. Reinstall the Agilent G1369A LAN card.
5. Connect the category 5 Ethernet cables as described in “Connecting the Ethernet Cables”
on page 28.
6. Reconnect the module to line power and turn on the module.
Switch #
678
Position
Off On Off
CAUTION To avoid an electric shock, always turn off an Agilent module and unplug
it from line power before you remove its cover.
Switch #
456
Position
Off On Off