Technical information
2
Setting Up the LAN Interface
Step 2: Setting Up the LAN Interface for a Telnet Session
8 Agilent 1200 Series Getting Connected Guide Thermo Scientific
3. If an Ethernet cable is connected to the Ethernet port, disconnect the cable.
4. Ensure that you are wearing ESD protection.
5. Remove the LAN card from the device (refer to the Agilent G1360A LAN Interface User
Manual).
The Agilent G1369A LAN card has a configuration switch that can be configured for
four initialization modes. The card is shipped with all eight switches set to the Off
position, which is the Bootp initialization mode (Figure 5).
Figure 5. Model G1369A LAN card
6. Set up the G1369A LAN card initialization switches as follows:
With these switch settings, the LAN card uses the factory default settings to enable a
TCP/IP connection to the LAN interface with a Telnet session. The default settings are
IP address 192.168.254.11 and subnet mask 255.255.0.0.
7. Reinsert the G1369A LAN card and power on the devices in the Agilent1200 LC Stack.
8. Using a category 5 Ethernet cable, connect the G1369A LAN card directly to the data
system computer.
CAUTION To prevent damage to an instrument, always use ESD protection when
handling electronic boards and components.
Switch #
456
Position
Off On On
Configuration switch
12345678
ON