Technical information

1
Checking the Firmware Versions of the Agilent Devices
Checking the Firmware Versions with the Agilent Instant Pilot
2 Agilent 1200 Series Getting Connected Guide Thermo Scientific
Checking the Firmware Versions with the Agilent Instant Pilot
You can use the Agilent Instant Pilot to check the firmware versions of the Agilent 1200 Series
modules to ensure their compatibility with the device drivers provided with LC Devices.
Y To check the firmware version of an Agilent 1200 Series module
1. Make sure that the Agilent 1200 Series modules are connected by CAN communication
cables, that all the modules are turned on, and that the Agilent 1200 Series Instant Pilot is
connected.
For instructions on how to connect the CAN cables, refer to the Agilent 1200 Series LC
reference manuals.
2. From the Welcome screen of the Instant Pilot, press the Details key (see Figure 1).
Figure 1. Welcome screen
The System Info screen appears. The System Info screen contains information about the
firmware versions of the configured system modules.
3. Check the firmware versions against those listed in Tabl e 1 on page 1.
Esc
Sequence
Status
Logbook
More
10:00
Select a button to continue
Display
Control
Details
Welcome
Quat Pump
Autosampler
Col Comp
Method
Agilent Technologies
7
8
9
456
12
3
0.~
i
Start Stop