User manual

G1369B LAN Interface User Manual 53
Getting Help 3
Troubleshooting
Error Messages
The error messages are shown in the LAN Interface Status Page on the Control
Module (G1323A/B) only, refer to “With Handheld Controller G1323B” on
page 43.
Figure 42 LAN Interface Status Page (complete)
If status "Error" shows up, possible error conditions are listed below.
Agilent Technologies G1369B
FW Revision : A.01.01
MAC Address : 0030d3060122
--------------------------------
Init Mode : Bootp
--------------------------------
TCP/IP Properties
- active -
IP Address : 0.0.0.0
Subnet Mask : not specified
Def. Gateway : not specified
- stored -
IP Address : 134.40.24.160
Subnet Mask : 255.255.248.0
Def. Gateway : 134.40.24.1
--------------------------------
TCP/IP Status : Error
Bootp timeout
--------------------------------
Controller : not connected
Possible reasons:
Bootp server not started or no
settings for this MAC address
found
Table 1 1 Error conditions
Error Description Action
Bootp timeout No reply on Bootp request received Start Bootp server and/or add settings
for the LAN Interface.
Bootp reply incomplete Bootp reply contained not all
information
Complete the neccessary information
Gateway in unreachable network Default Gateway does not match the
specified IP address and Subnet Mask
Correct the settings