User`s guide
71
Connecting Instruments to GPIB
Step 4: Configure GPIB Interface Cards
NOTE
After the system is configured, this screen may also display an
Edit VISA
Config...
button. Clicking this button allows you to manually configure the
interface as desired.
82350 GPIB Interface Card Configuration Parameters
SICL Interface
Name
Symbolic name that SICL uses to uniquely identify this GPIB interface. The
default Interface Name is gpib0. The SICL Interface Name must be a unique
string of alphanumeric characters, starting with a letter. Remember this value
and the Logical Unit number to properly address GPIB devices in your SICL
applications.
VISA Interface
Name
Symbolic name that VISA uses to uniquely identify this GPIB interface. The
default VISA Interface Name is GPIB0. The 82350 Interface Name for VISA must
begin with the string GPIB and have an integer appended to it, such as GPIB0,
GPIB1, GPIB2, etc. Remember this value to properly address GPIB devices in
your VISA applications.
Logical Unit
Number that SICL uses to uniquely identify this 82350 interface. The Logical Unit
number is an integer in the range of 0 - 10000. Remember this value and the
SICL Interface Name to properly address the GPIB interface in your SICL
applications.
Bus Address
address of this GPIB interface controller on the GPIB bus. It is usually 21 if the
GPIB interface is a System Controller or 20 if the GPIB interface is a non-System
Controller (see System Controller, following). These addresses are chosen by
convention but any address in the range 0 - 30, inclusive, may be used.
System
Controller
Determines if this interface controls which bus devices talk and which bus
devices listen. If several devices exist on a bus, be sure each has a unique GPIB
bus address and only one device is the System Controller (it is usually the device
installed in the computer). Each GPIB interface has its own independent bus.
Thus, each interface may be a System Controller as long as it is not chained
together with other GPIB interfaces. However, two or more System Controllers
on the same bus will cause the bus to be inoperative.