User`s guide

Agilent 4155C/4156C VXIplug&play Driver Users Guide, Edition 1 4-37
Sample Application Programs for Agilent VEE
Customizing Sample Programs
NOTE The modification example shown above changes the meaning of the wafer map
result display as shown below.
P Test results of all devices are within the allowable range.
F1 Device 1 test result is out of the allowable range.
F2 Device 2 test result is out of the allowable range.
F3 Device 3 test result is out of the allowable range.
F4 Test results of device 1 and 2 are out of the allowable range.
F5 Test results of device 2 and 3 are out of the allowable range.
F6 Test results of device 1 and 3 are out of the allowable range.
F7
Test results of all devices are out of the allowable range.
To Add a Measurement Parameter
If you want to add a measurement parameter, such as drain current Id, modify the
program as shown below. This example modifies sample2.vee for device 1 only
(this example does not modify the objects for device 2).
Adds the measurement function to the Measurement object.
Adds the object to set the dummy data to the Meas 4155 (Offline) object.
Adds the object to set the measurement source to the Vth Measurement object.
Adds the object to save the measured data to the Device 1 Measurement object.
Modifies the Show Result object and the Main panel display.
Figure 4-24 Id Measurement Setup
A
194
S
SUB
1
4
3
2
OUT
IN
SMU1
SMU2
SMU3
SMU4
4155/4156
E5250A
Cascade Summit series semi-auto prober
0.5V
3.0V
Device1
0V0V
327523
DG
GSD
SUB
Device2