Performance Tests Performance Test Instructions
54 Agilent 81560A, 81561A, 81566A, & 81567A Optical Attenuator Modules, First Edition
Relative Power Meter Uncertainty Test
Carry out the following Power Meter Uncertainty Test at 1550 nm or
another specified wavelength with single mode fibers using the
equipment listed in Table 2: Equipment required. To adapt for the
straight or angled contact versions of the attenuator use the
patchcords with appropriate connectors and the adequate connector
interfaces.
This test applies only to the Agilent 81566A and 815667A Attenuators
with integrated power meter.
NOTE Do not turn the laser off during the measurement. Clean all connectors
carefully before you start with the measurement !
1 Turn the instruments on and allow the devices to warm up (20..30
min).
2 Make sure that you perform this test in a temperature-controlled
environment with temperature fluctuations less than ±1°C.
3 Make sure that all your connectors are clean and undamaged.
4 Connect the equipment as shown below in Figure 17
Figure 17 Relative Power Meter Uncertainty Setup
5 Set both attenuators to the wavelength of the source and disable
them.
6 Set the attenuation [α] of 81561A reference attenuator to 55dB.
7 Set the attenuation [α] of the DUT attenuator to 0dB.
8 Set the averaging time of the DUT attenuator power meter to 1s.
9 Zero the reference and the attenuator power meter.