Performance Tests Performance Test Instructions
50 Agilent 81560A, 81561A, 81566A, & 81567A Optical Attenuator Modules, First Edition
Wavelength = Source Wavelength
•Range Mode = auto
8 Set the 11896A Polarization Controller scan rate = 4 and press
[AutoScan] to start Polarization Scrambling.
9 Press Measure at the power meter to start PDL Scanning.
10 After the measurement press [Analysis] and [More] to get the power
readings.
11 Note the results in the Test Record.
Polarization Dependent Loss (PDL) Test - Mueller
Method
Carry out the following PDL Test at 1550 nm with single mode fibers
using the equipment listed in Table 3, “Equipment Required,” on
page 41. To adapt for the straight or angled contact versions of the
attenuator use the patchcords with appropriate connectors and the
adequate connector interfaces. Use the operation mode "Attenuation
Set".
NOTE Perform this test with the plate-type 8169A Polarization Controller.
1 Turn the instruments on and allow the devices to warm up (20..30
min).
2 Make sure that all your connectors are clean and undamaged.
3 Connect the equipment as shown in Figure 15.
Figure 15 PDL Reference Setup
8163A
In
Out
8169
A