User`s guide

Performance Tests Power Tests
96 Agilent 81600B Tunable Laser Source Family, Fourth Edition
d Select <dB> as the power units,
e Set λ, the wavelength, to the same as the TLS module, as given in
step 3.
7 Switch on the High Power output of the TLS.
8 Note the power value displayed by the 81624A/B in the Test Record.
[For the 81600B #200, #160, #150, #140 and #130 use the Test Record
table “Power Linearity Output 2, High Power Upper Power Levels”.
For the 81600B #142 #003 use the Test Record table “Power Linearity
81600B #142 #003, High Power Upper Power Levels”.]
9 At the 81624A/B, select <Menu> then <Disp Ref>
10 Change the power setting of the TLS to the next value given in the Test
Record.
11 Note the (relative) power displayed by the 81624A/B as the “Measured
Relative Power from start”.
12 Calculate the "Power Linearity at current setting" as the sum of
"Measured Relative Power from start" and "Power Reduction from
start".
13 Repeat step 10 to step 12 for all power levels listed in the Test Record.
14 Determine the maximum value and the minimum value of the
calculated Power Linearity at the various settings and record them in
the Test Record as "Maximum Power Linearity at current setting", and
"Minimum Power Linearity at current setting", respectively.
15 Subtract the minimum power linearity value from the maximum power
linearity value and record the result as the Tota l Po we r Li ne ar ity.
[For example, refer to Ta b le .]
Test Procedure - High Power output, using the built-in attenuator
This test procedure is applicable to the:
It measures the power linearity of the module at the outputs indicated and
the built-in attenuator is used.
Agilent 81600B #200 (Output 2, High Power)
Agilent 81600B #160 (Output 2, High Power)
Agilent 81600B #150 (Output 2, High Power)
Agilent 81600B #140 (Output 2, High Power)
Agilent 81600B #130 (Output 2, High Power)
Agilent 81600B #142 #003