User`s guide
Performance Tests Power Tests
94 Agilent 81600B Tunable Laser Source Family, Fourth Edition
14
Determine the maximum value and the minimum value of the
calculated Power Linearity at the various settings and record them in
the Test Record as "Maximum Power Linearity at current setting", and
"Minimum Power Linearity at current setting", respectively.
15 Subtract the minimum power linearity value from the maximum power
linearity value and record the result as the Tota l Po we r Li ne ar ity.
Test Procedure - High Power output, no attenuation
This test procedure is applicable to the:
It measures the power linearity of the module at the outputs indicated.
Example: Agilent 81600B #200 Power Linearity, Output 1.
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity
at current setting
Start = REF + 2.0 dBm 0.00 dBm + 0.00 dBm = 0.00 dBm
+ 1.0 dBm - 1.02 dBm + 1.00 dBm = - 0.02 dBm
0.0 dBm - 1.92 dBm + 2.00 dBm = + 0.08 dBm
-1.0 dBm - 2.95 dBm + 3.00 dBm = + 0.05 dBm
-2.0 dBm - 4.07 dBm + 4.00 dBm = - 0.07 dBm
-3.0 dBm - 4.96 dBm + 5.00 dBm = + 0.04 dBm
-4.0 dBm - 5.97 dBm + 6.00 dBm = + 0.03 dBm
-5.0 dBm - 6.98 dBm + 7.00 dBm = + 0.02 dBm
-6.0 dBm - 7.97 dBm + 8.00 dBm = + 0.03 dBm
-7.0 dBm - 8.98 dBm + 9.00 dBm = + 0.02 dBm
Maximum Power Linearity at current setting: + 0.08 dBm
Minimum Power Linearity at current setting: - 0.07 dBm
Total Linearity
= Max. Power Linearity - Min. Power Linearity 0.15 dBpp
Agilent 81600B #200 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #160 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #150 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #140 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #130 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #142 #003, built-in attenuator not used.
Agilent 81600B #142
Agilent 81600B #132