User`s guide

Performance Tests Wavelength Accuracy
86 Agilent 81600B Tunable Laser Source Family, Fourth Edition
1
Move to the TLS channel of the Agilent 8164A/B Lightwave
Measurement System and press [Menu].
2 Set the menu parameters to:
3 Connect the fiber to the optical output.
If you are using the Agilent 81600B #200, #160, #150, #140 or #130,
connect the fiber to Output 2, the High Power output.
That is, set <Optical Output> to <High Power (2)>.
4 Set the wavelength and power of your TLS module to:
5 Switch on the TLS output.
6 Wait until the wavelength meter has settled then note the wavelength
displayed by the wavelength meter in the Test Record as the Initial
Setting, the reference wavelength “REF”.
7 Set the TLS module to any wavelength in its range. In the Test Record,
this is given in the “from wavelength” column.
Tunable Laser Channel Menu Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Automatic>
Modulation: <mod src><Off>
TLS module Wavelength [λ] Power [P]
Agilent 81600B #200 1440.000 nm - 1.00 dBm
Agilent 81600B #160 1495.000 nm - 1.00 dBm
Agilent 81600B #150 1450.000 nm - 1.00 dBm
Agilent 81600B #140 1370.200 nm - 3.00 dBm
Agilent 81600B #130 1260.000 nm - 3.00 dBm
Agilent 81600B #142 #003 1370.200 nm - 1.50 dBm
Agilent 81600B #142 1370.200 nm 0.00 dBm
Agilent 81600B #132 1260.000 nm 0.00 dBm
Water absorption lines For 81600B #140 and #142, some wavelengths
are set to odd values to avoid conflict with water absorption lines.
NOTE