User`s guide

Wavelength Accuracy Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 85
Figure 22 Wavelength Repeatability.
At the start of the test the TLS is set:
To its lowest specified wavelength,
To the highest power the TLS can deliver over the full wavelength range,
Such that any modulation is off,
At the end of the test, the TLS is set to its maximum specified wavelength.
Test Procedure
This test procedure is applicable to the:
It measures the wavelength repeatability of the module at the outputs
indicated.
Attenuate the Power Output from the TLS Reduce the output power
delivered by the TLS to a level compatible with the capabilities of the
wavelength meter.
Use the TLS module’s built-in attenuator, or an external attenuator.
NOTE
Agilent 81600B #200 (Output 2, High Power)
Agilent 81600B #160 (Output 2, High Power)
Agilent 81600B #150 (Output 2, High Power)
Agilent 81600B #140 (Output 2, High Power)
Agilent 81600B #142 #003
Agilent 81600B #142
Agilent 81600B #130 (Output 2, High Power)
Agilent 81600B #132