User`s guide

Wavelength Accuracy Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 83
1 Move to the TLS channel of the Agilent 8164A/B Lightwave
Measurement System and press [Menu].
2 Set the menu parameters to:
3 Connect the fiber to the optical output.
If you are using the Agilent 81600B #200, #160, #150 or #140, connect
the fiber to Output 2, the High Power output.
That is, set <Optical Output> to <High Power (2)>.
4 Set the inital wavelength and power of your TLS to:
5 Switch on the TLS output.
6 Wait until the wavelength meter has settled then note the wavelength
displayed by the wavelength meter in the test record.
7 Set the TLS module to the next wavelength given in the Test record.
8 Repeat step 6 and step 7 to the maximum wavelength value for the TLS
module:
Tunable Laser Channel Menu Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Automatic>
Modulation: <mod src><Off>
TLS module Wavelength [λ] Power [P]
Agilent 81600B #200 1440.000 nm - 1.00 dBm
Agilent 81600B #160 1495.000 nm - 1.00 dBm
Agilent 81600B #150 1450.000 nm - 1.00 dBm
Agilent 81600B #140 1370.200 nm - 3.00 dBm
Agilent 81600B #130 1260.000 nm - 3.00 dBm
Agilent 81600B #142 #003 1370.200 nm - 1.50 dBm
Agilent 81600B #142 1370.200 nm 0.00 dBm
Agilent 81600B #132 1260.000 nm 0.00 dBm
Water absorption lines For 81600B #140 and #142, some wavelengths
are set to odd values to avoid conflict with water absorption lines.
NOTE