User`s guide
Performance Tests Wavelength Accuracy
80 Agilent 81600B Tunable Laser Source Family, Fourth Edition
5
Switch on the TLS output.
6 Wait until the wavelength meter has settled then note the wavelength
displayed by the wavelength meter in the Test Record.
7 Set the TLS module to the next wavelength given in the Test Record.
8 Repeat step 6 and step 7 to the maximum wavelength value for the TLS
module:
9 Repeat step 4 to step 8 another four times.
10 Select the maximum and minum deviations from each repetition of the
measurements, and note these values in the Test Record.
11 Determine the Relative Wavelength Accuracy and Summary of all
Repetitions.
a Take the Largest Maximum Deviation and note it in the Test Record,
b Take the Smallest Minimum Deviation and note it in the Test Record.
12 Determine, and note in the Test Record, the
Relative Wavelength Accuracy, which is the
Water absorption lines For 81600B #130, #132, #140 and #142, some
wavelengths are set to odd values to avoid conflict with water absorption
lines.
NOTE
TLS module Maximum Wavelength Value
Agilent 81600B #200 1640 nm
Agilent 81600B #160 1640 nm
Agilent 81600B #150 1590 nm
Agilent 81600B #140 1495 nm
Agilent 81600B #130 1375 nm
Agilent 81600B #142 #003 1495 nm
Agilent 81600B #142 1495 nm
Agilent 81600B #132 1375 nm
Determining the Maximum and the Minimum Deviations
•The Largest Maximum Deviation is the most positive value (or the least
negative, if all values are negative).
•The Smallest Minimum Deviation is the most negative value (or the least
positive, if all values are positive).
NOTE