User`s guide
Required Test Equipment Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 73
Required Test Equipment
The equipment required for the performance test is listed in Ta ble 6 . Any
equipment that satisfies the critical specifications of the equipment given
in Ta bl e 6 may be substituted for the recommended models.
Table 6 Equipment Required
Instrument/Accessory Description #071 #072
Agilent 86142B
a
Optical Spectrum analyser 1 1
Agilent 8164A/B Lightwave Measurement System 1 1
Agilent 86122A
b
MultiWavelength meter 1 1
Agilent 81618A or 81619A Optical Head Interface Module 1 1
Agilent 81624A/B #C01
c
Standard Optical Head 1 1
Agilent 81634A/B Power Sensor 1 1
Agilent 81000SA DIN 47256/4108 Connector Adapter 1 1
Agilent 81000AI HMS-10 Connector Interface 1
Agilent 81000SI DIN 47256/4108 Connector Interface 2 2
Agilent 81000FI FC/PC Connector Interface 1 1
Agilent 81101AC Diamond HMS-10 — Diamond HMS-10 Patchcord 1
Agilent 81101PC Diamond HMS-10 — Agilent FC/PC Patchcord 1
Agilent 81113PC Diamond HMS-10 — Agilent FC/Super PC Patchcord 1 1
Agilent 81113SC Diamond HMS-10 — Agilent DIN 47256/4108 Patchcord 1
1005-0255 Adapter Din - DIN 1
N/A Fiber Bragg Grating
d
11
Equipment for optional tests
81636B
e
Fast Power Meter 1 1
N/A Wavelength Reference Unit (Faby-Perot etalon)
f
11
N/A Wavelength Reference Unit (Michelson Interferometer) - optional 1 1
a
You can use the HP 71452B or HP 71450A #100 instead of the Agilent 86142B.
b
You can use the Burleigh WA-1500 Wavemeter instead of the Agilent 86120C.
c
You can use the 81623A/B #C01 instead of the 81624A/B #C01
d
For the 81600B #200, #160 and #150 approximately 1520 nm, 2 nm at 3 dB
For the 81600B #140 approximately 1407 nm, 2 nm at 3 dB
e
You can use the 81637B instead of the 81636B. Required characteristic: Sample rate ≥ 40 kHz
f
Required characteristics:
- Optical length 9.35 ± 0.08 mm at 1510 nm
- Reflectivity 50 ± 2 %
- Wavelength range 1250 nm - 1650 nm
- Birefringency DIN 3140-6, that is 20 nm/cm or 2*10
-6
- Linear polarizer with AR-coating at FP-etalon input (~ 30 dB extinction ratio, aligned with the principle state of polarization)
- Temperature Dependency: drift < 0.1 pm over the duration of the test (~ 15 min.).
A reasonable target temperature coefficient is < 0.3 pm/K, which typically requires active temperature regulation.
- Insertion Loss (minimum value over the specified wavelength range) < 3.5 dB
Fiber connections: Angled PM fiber at input (requires DUT independent patchcord) and angled SM fiber at output.