User`s guide
Specifications Conditions
66 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Conditions
Storage Temperature: -40° C to +70° C
Operating Temperature: +10°
C to +35° C
Humidity: < 80% R.H. at 10° C to +35° C non-condensing.
Specifications apply for wavelengths not equal to any water absorption
line.
Warm-up time: 1 h ; immediate operation after bootup
Output Power: Specifications are valid at the following output power levels:
81600B option 200, 160, and 150:
≥ -7 dBm (for Output 1)
≥ -1 dBm (for Output 2, attenuation 0 dB).
81600B option 140:
≥ -13 dBm (for Output 1)
≥ -3 dBm (for Output 2).
81600B option 130:
≥ -13 dBm (for Output 1)
≥ -3 dBm (for Output 2).
81600B option 142:
≥ -3 dBm
≥ -4.5 dBm (with option 003).
81600B option 132:
≥ 0 dBm
Continuous sweep mode: Specifications are valid for mode-hop free sweeping.
Maximum 50 nm at constant output power levels as follows:
81600B option 200:
1475 nm - 1620 nm
≥ -2 dBm (for Output 1)
≥ +4 dBm (for Output 2).