User`s guide
Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 237
Agilent 81600B #132 Performance Test Page 7 of 11
Model Agilent 81600B #132 Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty:
• Using 81624A/B #C01 ±2.8%
• Using 81623A/B #C01 ±3.5% (up to 8 dBm)
Power Linearity
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.2 dBpp
Measurement Uncertainty Using 81624A/B #C01 ± 0.02 dB
Wavelength Setting Power Measured Lower Test Limit
1260.000 nm dBm 0.00 dBm
1270.000 nm dBm + 3.00 dBm
1280.000 nm dBm + 3.00 dBm
1290.000 nm dBm + 7.00 dBm
1300.000 nm dBm + 7.00 dBm
1310.000 nm dBm + 7.00 dBm
1320.000 nm dBm + 7.00 dBm
1330.000 nm dBm + 7.00 dBm
1340.000 nm dBm + 7.00 dBm
1350.000 nm dBm + 7.00 dBm
1360.000 nm dBm + 7.00 dBm
1369.500 nm dBm + 7.00 dBm
1375.000 nm dBm + 3.00 dBm
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF + 7.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
+ 6.0dBm dB + 1.00dB = dB
+ 5.0dBm dB + 2.00dB = dB
+ 4.0dBm dB + 3.00dB = dB
+ 3.0dBm dB + 4.00dB = dB
+ 2.0dBm dB + 5.00dB = dB
+ 1.0dBm dB + 6.00dB = dB
0.0 dBm dB + 7.00 dB = dB