User`s guide
Performance Tests Test Records
224 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #142 Performance Test Page 8 of 13
Model Agilent 81600B #142 Tunable Laser Report No. ________ Date_______
Power Linearity Output 1, Low SSE Option #003
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.2 dBpp
Measurement Uncertainty Using 81624A/B #C01 ± 0.02 dB
Using 81623A/B #C01 ± 0.045 dB
Power Linearity Output 1, Low SSE Option #003 Upper Power Levels
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.6 dBpp
Measurement Uncertainty Using 81624A/B #C01 ± 0.05 dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF + 7.5 dBm 0.00 dB + 0.00 dB = 0.00 dB
+ 7.0dBm dB + 0.50dB = dB
+ 6.0dBm dB + 1.50dB = dB
+ 5.0dBm dB + 2.50dB = dB
+ 4.0dBm dB + 3.50dB = dB
+ 3.0dBm dB + 4.50dB = dB
+ 2.0dBm dB + 5.50dB = dB
+ 1.0dBm dB + 6.50dB = dB
0.0 dBm dB + 7.50 dB = dB
- 1.0 dBm dB + 8.50 dB = dB
- 2.0 dBm dB + 8.50 dB = dB
- 3.0 dBm dB + 9.50 dB = dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF + 6.0 dBm 0.00 dB + 0.00 dB = dB
+ 5.0dBm dB + 1.00dB = dB
+ 4.0dBm dB + 2.00dB = dB
+ 3.0dBm dB + 3.00dB = dB
+ 2.0dBm dB + 4.00dB = dB
+ 1.0dBm dB + 5.00dB = dB
0.0 dBm dB + 6.00 dB = dB
- 1.0 dBm dB + 7.00 dB = dB
- 2.0 dBm dB + 8.00 dB = dB
- 3.0 dBm dB + 9.00 dB = dB
- 4.0 dBm dB 10.00 dB dB
- 4.5 dBm dB 10.50 dB dB