User`s guide

Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 223
Agilent 81600B #142 Performance Test Page 7 of 13
Model Agilent 81600B #142 Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty:
Using 81624A/B #C01 ±2.8%
Using 81623A/B #C01 ±3.5% (up to 8 dBm)
Output 1 Output 2
Wavelength Setting Power Measured Lower Test Limit Power Measured Lower Test Limit
1370.200 nm dBm 0.00 dBm dBm - 1.50 dBm
1380.000 nm dBm 0.00 dBm dBm - 1.50 dBm
1390.200 nm dBm 0.00 dBm dBm - 1.50 dBm
1400.200 nm dBm 0.00 dBm dBm - 1.50 dBm
1410.000 nm dBm 0.00 dBm dBm - 1.50 dBm
1420.200 nm dBm + 5.00 dBm dBm + 3.50 dBm
1430.200 nm dBm + 7.50 dBm dBm + 6.00 dBm
1440.000 nm dBm + 7.50 dBm dBm + 6.00 dBm
1450.000 nm dBm + 7.50 dBm dBm + 6.00 dBm
1460.000 nm dBm + 7.50 dBm dBm + 6.00 dBm
1470.000 nm dBm + 7.50 dBm dBm + 6.00 dBm
1480.000 nm dBm + 7.50 dBm dBm + 6.00 dBm
1495.000 nm dBm 0.00 dBm dBm - 1.50 dBm