User`s guide

Performance Tests Test Records
222 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #142 Performance Test Page 6 of 13
Model Agilent 81600B #142 Tunable Laser Report No. ________ Date_______
Wavelength Repeatability
Measurement Uncertainty: ± 0.1 pm
Repeatability of
1370.200 nm (= reference)
Measurement Result
Repeatability of
1430.200 nm (= reference)
Measurement Result
Initial Setting REF = nm Initial Setting REF = nm
from 1390.200 nm to REF nm from 1370.200 nm to REF nm
from 1420.200 nm to REF nm from 1390.200 nm to REF nm
from 1450.000 nm to REF nm from 1450.000 nm to REF nm
from 1495.000 nm to REF nm from 1495.000 nm to REF nm
largest measured wavelength nm largest measured wavelength nm
smallest measured
wavelength
nm smallest measured
wavelength
nm
Wavelength Repeatability nm Wavelength Repeatability nm
= largest measured wavelength - smallest measured
wavelength
= largest measured wavelength - smallest measured
wavelength
Upper Test Limit 0.0016 nm Upper Test Limit 0.0016 nm
Performance Characteristic 0.0010 nm typical Performance Characteristic 0.0010 nm typical
Repeatability of
1495.000 nm (= reference)
Measurement
Result
Initial Setting REF =
nm
from 1370.200 nm to REF nm
from 1390.200 nm to REF nm
from 1420.200 nm to REF nm
from 1450.000 nm to REF nm
largest measured wavelength nm
largest measured wavelength nm
smallest measured
wavelength
nm
Wavelength Repeatability nm
= largest measured wavelength - smallest measured
wavelength
Upper Test Limit 0.0016 nm
Performance Characteristic 0.0010 nm typical