User`s guide
Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 201
Agilent 81600B #140 Performance Test Page 11 of 14
Model Agilent 81600B #140 Tunable Laser Report No. ________ Date_______
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission - Output 2, High Power
Measurement Uncertainty: ± 0.20 dB
Low SSE Output 1 High Power Output 2
Att = 0 dB
Maximum Deviation dB dB
Minimum Deviation dB dB
Power Stability
1
dB dB
Upper Test Limit 0.02 dBpp 0.02 dBpp
Measurement Uncertainty ±0.005 dB ±0.005 dB
Wavelength Output Power Results Lower Test Limit
1370.000 nm - 3.00 dBm dB 30 dB(typ. 35 dB)
1380.000 nm - 3.00 dBm dB 30 dB(typ. 35 dB)
1390.000 nm - 3.00 dBm dB 30 dB(typ. 35 dB)
1400.000 nm - 3.00 dBm dB 30 dB(typ. 35 dB)
1410.000 nm - 3.00 dBm dB 30 dB(typ. 35 dB)
1420.000 nm + 3.00 dBm dB 40 dB
1430.000 nm + 5.00 dBm dB 42 dB
1440.000 nm + 5.00 dBm dB 42 dB
1450.000 nm + 5.00 dBm dB 42 dB
1460.000 nm + 5.00 dBm dB 42 dB
1470.000 nm + 5.00 dBm dB 42 dB
1480.000 nm + 5.00 dBm dB 42 dB
1495.000 nm - 3.00 dBm dB 30 dB(typ. 35 dB)