User`s guide
Performance Tests Test Records
200 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #140 Performance Test Page 10 of 14
Model Agilent 81600B #140 Tunable Laser Report No. ________ Date_______
Power Flatness
Wavelength
Low SSE Output 1 High Power Output 2
P = −13.00 dBm
P = −3dBm
ATT = 0 dB
P = −3dBm
ATT = 57 dB
Power Deviation Power Deviation Power Deviation
Start = REF 1420.200 nm 0.00 dB 0.00 dB 0.00 dB
1425.000 nm dB dB dB
1430.200 nm dB dB dB
1435.000 nm dB dB dB
1440.000 nm dB dB dB
1445.000 nm dB dB dB
1450.000 nm dB dB dB
1455.000 nm dB dB dB
1460.000 nm dB dB dB
1465.000 nm dB dB dB
1470.000 nm dB dB dB
1480.000 nm dB dB dB
1495.000 nm dB dB dB
Maximum deviation dB dB dB
Minimum deviation dB dB dB
Flatness = Maximum − Minimum Deviation dB dB dB
Upper Test Limit 0.40 dBpp 0.60 dBpp 0.60 dBpp
Measurement Uncertainty:
Using 81624A/B #C01 ± 1.6% ± 1.6% ± 1.6%
Using 81623A/B #C01 ± 1.6% ± 1.6% ± 1.6%