User`s guide
Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 197
Agilent 81600B #140 Performance Test Page 7 of 14
Model Agilent 81600B #140 Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty:
• Using 81624A/B #C01 ±2.8%
• Using 81623A/B #C01 ±3.5% (up to 8 dBm)
Output 1 Output 2
Wavelength Setting Power Measured Lower Test Limit Power Measured Lower Test Limit
1370.200 nm dBm - 13.00 dBm dBm - 3.00 dBm
1380.000 nm dBm - 13.00 dBm dBm - 3.00 dBm
1390.200 nm dBm - 13.00 dBm dBm - 3.00 dBm
1400.200 nm dBm - 13.00 dBm dBm - 3.00 dBm
1410.000 nm dBm - 13.00 dBm dBm - 3.00 dBm
1420.200 nm dBm - 7.00 dBm dBm + 3.00 dBm
1430.200 nm dBm - 5.00 dBm dBm + 5.00 dBm
1440.000 nm dBm - 5.00 dBm dBm + 5.00 dBm
1450.000 nm dBm - 5.00 dBm dBm + 5.00 dBm
1460.000 nm dBm - 5.00 dBm dBm + 5.00 dBm
1470.000 nm dBm - 5.00 dBm dBm + 5.00 dBm
1480.000 nm dBm - 5.00 dBm dBm + 5.00 dBm
1495.000 nm dBm - 13.00 dBm dBm - 3.00 dBm