User`s guide

Performance Tests Test Records
190 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #150 Performance Test Page 14 of 14
Model Agilent 81600B #150 Tunable Laser Report No. ________ Date_______
Optional Tests:
Dynamic Wavelength Accuracy and Repeatability
Sweep speed 5 nm/s 40 nm/s 80 nm/s
Scan # 123451234512345
Absolute static wavelength
accuracy (from page 4)
Relative static wavelength
accuracy (from page 4)
∆λ
REL
(n)
∆λ
OFFSET
(n)
Dynamic relative wavelength
accuracy per scan, R(n)
Dynamic absolute wavelength
accuracy per scan, A(n)
Average dynamic absolute wave-
length accuracy, A
AVG
__________ pm __________ pm __________ pm
Testlimits (static + dynamic add-
on)
± 10.4 pm ± 11 pm ± 12.5 pm
Average dynamic relative wave-
length accuracy, R
AVG
__________ pm __________ pm __________ pm
Testlimits (static + dynamic add-
on)
± 5.4 pm ± 5.8 pm ± 7 pm
Sweep speed 20 nm/s 40 nm/s 80 nm/s
Dynamic Wavelength Repeatabil-
ity, REP
peak to peak
__________ pm __________ pm __________ pm
Test Limit (peak to peak) 0.6 pm 0.8 pm 1.4 pm