Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 181
Agilent 81600B #150 Performance Test Page 5 of 14
Model Agilent 81600B #150 Tunable Laser Report No. ________ Date_______
Mode Hop Free Tuning
1
Wavelength Deviation = Wavelength Measured - Wavelength Setting
Mode Hop Free Tuning Result (= Largest value of either the Maximum or Minimum Deviation)
Mode Hop Free Tuning Result: _______ nm
Te s t L i mi t : ±0.025 nm
Measurement Uncertainty: ±0.2 pm
Wavelength Setting Wavelength Measured Wavelength Deviation
1
1450.000 nm nm nm
1451.000 nm nm nm
1452.000 nm nm nm
1453.000 nm nm nm
1454.000 nm nm nm
1455.000 nm nm nm
1456.000 nm nm nm
1457.000 nm nm nm
1458.000 nm nm nm
1459.000 nm nm nm
1460.000 nm nm nm
1580.000 nm nm nm
1581.000 nm nm nm
1582.000 nm nm nm
1583.000 nm nm nm
1584.000 nm nm nm
1585.000 nm nm nm
1586.000 nm nm nm
1587.000 nm nm nm
1588.000 nm nm nm
1589.000 nm nm nm
1590.000 nm nm nm
Maximum Deviation: nm
Minimum Deviation: nm