User`s guide

Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 173
Agilent 81600B #160 Performance Test Page 11 of 14
Model Agilent 81600B #160 Tunable Laser Report No. ________ Date_______
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission - Output 2, High Power
Measurement Uncertainty: ± 0.20 dB
Low SSE Output 1 High Power Output 2
Att = 0 dB
Maximum Deviation dB dB
Minimum Deviation dB dB
Power Stability
1
dB dB
Upper Test Limit 0.02 dBpp 0.02 dBpp
Measurement Uncertainty ±0.005 dB ±0.005 dB
Wavelength Output Power Results Lower Test Limit
1495.000 nm − 1.00 dBm dB 37 dB
1510.000 nm + 3.00 dBm dB 42 dB
1520.000 nm + 5.00 dBm dB 45 dB
1530.000 nm + 5.00 dBm dB 45 dB
1540.000 nm + 5.00 dBm dB 45 dB
1550.000 nm + 5.00 dBm dB 45 dB
1560.000 nm + 5.00 dBm dB 45 dB
1570.000 nm + 5.00 dBm dB 45 dB
1580.000 nm + 5.00 dBm dB 45 dB
1590.000 nm + 5.00 dBm dB 45 dB
1600.000 nm + 5.00 dBm dB 45 dB
1610.000 nm + 5.00 dBm dB 45 dB
1620.000 nm + 3.00 dBm dB 42 dB
1630.000 nm − 1.00 dBm dB 37 dB
1640.000 nm − 1.00 dBm dB 37 dB