User`s guide

Performance Tests Test Records
170 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #160 Performance Test Page 8 of 14
Model Agilent 81600B #160 Tunable Laser Report No. ________ Date_______
Power Linearity Output 1, Low SSE
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.2 dBpp
Measurement Uncertainty ± 0.05 dB
Power Linearity Output 2, High Power Upper Power Levels
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.2 dBpp
Measurement Uncertainty ± 0.05 dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF 4.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
5.0 dBm dB + 1.00 dB = dB
6.0 dBm dB + 2.00 dB = dB
7.0 dBm dB + 3.00 dB = dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity
at current setting
Start = REF + 5.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
+ 4.0 dBm dB + 1.00 dB = dB
+ 3.0 dBm dB + 2.00 dB = dB
+ 2.0 dBm dB + 3.00 dB = dB
+ 1.0 dBm dB + 4.00 dB = dB
0.0 dBm dB + 5.00 dB = dB
1.0 dBm dB + 6.00 dB = dB