User`s guide

Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 161
Agilent 81600B #200 Performance Test Page 13 of 14
Model Agilent 81600B #200 Tunable Laser Report No. ________ Date_______
Signal-to-Total-Source Spontaneous Emission - Output 1, Low SSE
Center Wavelength of Fiber Bragg Grating: TLS_λ
0
= ________ nm
OSA_λ
0
= ________ nm
Transmission Band Limits: OSA_λ
1
= ________ nm
OSA_λ
2
= ________ nm
OSA_λ
3
= ________ nm
OSA_λ
4
= ________ nm
* (for TLS_l
0
= 1520 nm - 1610 nm)
Measurement Uncertainty: ± 2.0 dB
Optional Test: Signal-to-Total-Source Spontaneous Emission - Output 2, High Power
* (for TLS_l
0
= 1520 nm - 1610 nm)
Measurement Uncertainty: ± 2.00 dB
Output 1, Low SSE
OSA_noise _________ pW
Sum of all SSE power levels in lower transmission band _________ pW
Sum of all SSE power levels in upper transmission band _________ pW
power_trans = Sum of all SSE power levels in transmission
bands
_________ pW
power_att _________ pW
power_total_noise= power_trans + power_att _________ pW
peak_power _________ pW
Measurement Result - Total SSE _________ dB
Lower Test Limit: 65 dB*
Total SSE 10
peak power
power
total noise OSA noise
--- ----- ---- ------ ----- ---- ------ ---- ----- -------- --------- ---- ------- ---- ---- ------- ---- ---- --log×=
Output 2, High Power
OSA_noise _________ pW
SSE_power_λTLS_max _________ pW
Power_total_noise = OSA_noise + SSE_power_λTLS_max _________ pW
Peak_power _________ pW
Measurement Result - Total SSE _________ dB
Te s t l i m i t 27 dB *
Performance Characteristic 30 dB typical
Total SSE 10
peak power
power
total SSE
---- ---- ----- ------ ---- ----- ------ ---- ----- --------log×=