User`s guide

Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 159
Agilent 81600B #200 Performance Test Page 11 of 14
Model Agilent 81600B #200 Tunable Laser Report No. ________ Date_______
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission - Output 2, High Power
Measurement Uncertainty: ± 0.20 dB
Low SSE Output 1 High Power Output 2
Att = 0 dB
Maximum Deviation dB dB
Minimum Deviation dB dB
Power Stability
1
dB dB
Upper Test Limit 0.02 dBpp 0.02 dBpp
Measurement Uncertainty ±0.005 dB ±0.005 dB
Wavelength Output Power Results Lower Test Limit
1440.000 nm 1.00 dBm dB 37 dB
1450.000 nm 1.00 dBm dB 37 dB
1460.000 nm 1.00 dBm dB 37 dB
1470.000 nm 1.00 dBm dB 37 dB
1475.000 nm + 4.00 dBm dB 43 dB
1480.000 nm + 4.00 dBm dB 43 dB
1490.000 nm + 4.00 dBm dB 43 dB
1500.000 nm + 4.00 dBm dB 43 dB
1510.000 nm + 4.00 dBm dB 43 dB
1520.000 nm + 8.00 dBm dB 48 dB
1530.000 nm + 8.00 dBm dB 48 dB
1540.000 nm + 8.00 dBm dB 48 dB
1550.000 nm + 8.00 dBm dB 48 dB
1560.000 nm + 8.00 dBm dB 48 dB
1570.000 nm + 8.00 dBm dB 48 dB
1580.000 nm + 8.00 dBm dB 48 dB
1590.000 nm + 8.00 dBm dB 48 dB
1600.000 nm + 8.00 dBm dB 48 dB
1610.000 nm + 8.00 dBm dB 48 dB
1620.000 nm + 4.00 dBm dB 43 dB
1625.000 nm + 4.00 dBm dB 43 dB
1630.000 nm − 1.00 dBm dB 37 dB
1640.000 nm − 1.00 dBm dB 37 dB