User`s guide
Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 155
Agilent 81600B #200 Performance Test Page 8 of 14
Model Agilent 81600B #200 Tunable Laser Report No. ________ Date_______
Power Linearity Output 1, Low SSE
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.2 dBpp
Measurement Uncertainty Using 81624A/B #C01 ± 0.02 dB
Using 81623A/B #C01 ± 0.025 dB
Power Linearity Output 2, High Power Upper Power Levels
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF + 2.0dBm 0.00dB + 0.00dB = 0.00dB
+ 1.0 dBm dB + 1.00 dB = dB
0.0 dBm dB + 2.00 dB = dB
− 1.0 dBm dB + 3.00 dB = dB
− 2.0 dBm dB + 4.00 dB = dB
− 3.0 dBm dB + 5.00 dB = dB
− 4.0 dBm dB + 6.00 dB = dB
− 5.0 dBm dB + 7.00 dB = dB
− 6.0 dBm dB + 8.00 dB = dB
− 7.0 dBm dB + 9.00 dB = dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity
at current setting
Start = REF + 8.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
+ 7.0 dBm dB + 1.00 dB = dB
+ 6.0 dBm dB + 2.00 dB = dB
+ 5.0 dBm dB + 3.00 dB = dB
+ 4.0 dBm dB + 4.00 dB = dB
+ 3.0 dBm dB + 5.00 dB = dB
+ 2.0 dBm dB + 6.00 dB = dB
+ 1.0 dBm dB + 7.00 dB = dB
0.0 dBm dB + 8.00 dB = dB
− 1.0 dBm dB + 9.00 dB = dB