User`s guide
Performance Tests Test Records
154 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #200 Performance Test Page 7 of 14
Model Agilent 81600B #200 Tunable Laser Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty:
• Using 81624A/B #C01 ±2.8%
• Using 81623A/B #C01 ±3.5% (up to 8 dBm)
Output 1 Output 2
Wavelength Setting Power Measured Lower Test Limit Power Measured Lower Test Limit
1440.000 nm dBm − 7.00 dBm dBm − 1.00 dBm
1450.000 nm dBm − 7.00 dBm dBm − 1.00 dBm
1460.000 nm dBm − 7.00 dBm dBm − 1.00 dBm
1470.000 nm dBm − 7.00 dBm dBm − 1.00 dBm
1475.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1480.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1490.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1500.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1510.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1520.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1530.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1540.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1550.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1560.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1570.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1580.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1590.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1600.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1610.000 nm dBm + 2.00 dBm dBm + 8.00 dBm
1620.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1625.000 nm dBm − 2.00 dBm dBm + 4.00 dBm
1630.000 nm dBm − 7.00 dBm dBm − 1.00 dBm
1640.000 nm dBm − 7.00 dBm dBm − 1.00 dBm