User`s guide
Performance Tests Test Records
152 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #200 Performance Test Page 5 of 14
Model Agilent 81600B #200 Tunable Laser Report No. ________ Date_______
Mode Hop Free Tuning
1
Wavelength Deviation = Wavelength Measured - Wavelength Setting
Mode Hop Free Tuning Result (= Largest value of either the Maximum or Minimum Deviation)
Mode Hop Free Tuning Result: _______ nm
Test Limit: ±0.025 nm
Measurement Uncertainty: ±0.2 pm
Wavelength Setting Wavelength Measured Wavelength Deviation
1
1440.000 nm nm nm
1441.000 nm nm nm
1442.000 nm nm nm
1443.000 nm nm nm
1444.000 nm nm nm
1445.000 nm nm nm
1446.000 nm nm nm
1447.000 nm nm nm
1448.000 nm nm nm
1449.000 nm nm nm
1450.000 nm nm nm
1630.000 nm nm nm
1631.000 nm nm nm
1632.000 nm nm nm
1633.000 nm nm nm
1634.000 nm nm nm
1635.000 nm nm nm
1636.000 nm nm nm
1637.000 nm nm nm
1638.000 nm nm nm
1639.000 nm nm nm
1640.000 nm nm nm
Maximum Deviation: nm
Minimum Deviation: nm